Please use this identifier to cite or link to this item:
Title: Simultaneous detection of five different 2-hydroxyethyl-DNA adducts formed by ethylene oxide exposure, using a high-performance liquid chromatography/electrospray ionisation tandem mass spectrometry assay.
Authors: Tompkins, EM
Jones, DJ
Lamb, JH
Marsden, DA
Farmer, PB
Brown, K
First Published: 2008
Citation: RAPID COMMUN MASS SPECTROM, 2008, 22 (1), pp. 19-28
Abstract: A method has been developed for the simultaneous detection and quantitation of five different 2-hydroxyethyl-DNA (HE-DNA) adducts that could be formed as a result of exposure to ethylene oxide (EO). In addition to the major N7-HE-guanine (N7-HEG) adducts this assay can also measure the less prevalent but potentially more biologically significant N1-HE-2'-deoxyadenosine (N1-HEdA), O(6)-HE-2'-deoxyguanosine (O(6)-HEdG), N(6)-HE-2'-deoxyadenosine (N(6)-HEdA) and N3-HE-2'-deoxyuridine adducts (N3-HEdU). The method involves the isolation of HE adducts from the unmodified nucleosides by either neutral thermal hydrolysis or enzymatic digestion, followed by high-performance liquid chromatographic (HPLC) purification, before detection and quantification by liquid chromatography tandem mass spectrometry (LC/MS/MS) using selective reaction monitoring (SRM). The limits of detection were in the range 0.5-25 fmol for each individual adduct, making this one of the most sensitive assays available for the detection of N7-HEG. To illustrate the possible applications of the assay, it has been employed in the measurement of endogenous/background and EO-induced HE adducts in a variety of DNA samples.
DOI Link: 10.1002/rcm.3328
ISSN: 0951-4198
Type: Journal Article
Appears in Collections:Published Articles, Dept. of Cancer Studies and Molecular Medicine

Files in This Item:
There are no files associated with this item.

Items in LRA are protected by copyright, with all rights reserved, unless otherwise indicated.