Please use this identifier to cite or link to this item:
Title: Evaluation of XRI-UNO CdTe detector for nuclear medical imaging
Authors: Jambi, Layal K.
Lees, John Ernest
Bugby, S. L.
Tipper, S.
Alqahtani, M. S.
Perkins, A. C.
First Published: 22-Jun-2015
Citation: Journal of Instrumentation, 2015, 10, pp. ?-? (15)
Abstract: Over the last two decades advances in semiconductor detector technology have reached the point where they are sufficiently sensitive to become an alternative to scintillators for high energy gamma ray detection for application in fields such as medical imaging. This paper assessed the Cadmium-Telluride (CdTe) XRI-UNO semiconductor detector produced by X-RAY Imatek for photon energies of interest in nuclear imaging. The XRI-UNO detector was found to have an intrinsic spatial resolution of <0.5mm and a high incident count rate capability up to at least 1680cps. The system spatial resolution, uniformity and sensitivity characteristics are also reported.
DOI Link: 10.1088/1748-0221/10/06/P06012
eISSN: 1748-0221
Version: Publisher Version
Status: Peer-reviewed
Type: Journal Article
Rights: Copyright © 2015, the authors. Content from this work may be used under the terms of the Creative Commons Attribution 3.0 License. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.
Appears in Collections:Published Articles, Dept. of Physics and Astronomy

Files in This Item:
File Description SizeFormat 
1748-0221_10_06_P06012.pdfPublished (publisher PDF)2.27 MBAdobe PDFView/Open