Please use this identifier to cite or link to this item: http://hdl.handle.net/2381/38642
Title: A comparison of CsI:Tl and GOS in a scintillator-CCD detector for nuclear medicine imaging
Authors: Bugby, S. L.
Jambi, L. K.
Lees, J. E.
First Published: 22-Sep-2016
Publisher: IOP Publishing
Citation: Journal of Instrumentation, 2016, 11 P09009
Abstract: A number of portable gamma cameras for medical imaging use scintillator-CCD based detectors. This paper compares the performance of a scintillator-CCD based portable gamma camera with either a columnar CsI:Tl or a pixelated GOS scintillator installed. The CsI:Tl scintillator has a sensitivity of 40% at 140.5 keV compared to 54% with the GOS scintillator. The intrinsic spatial resolution of the pixelated GOS detector was 1.09 mm, over 4 times poorer than for CsI:Tl. Count rate capability was also found to be significantly lower when the GOS scintillator was used. The uniformity was comparable for both scintillators.
DOI Link: 10.1088/1748-0221/11/09/P09009
eISSN: 1748-0221
Links: http://iopscience.iop.org/article/10.1088/1748-0221/11/09/P09009/meta
http://hdl.handle.net/2381/38642
Version: Publisher Version
Status: Peer-reviewed
Type: Journal Article
Rights: Published under the terms of the Creative Commons Attribution 3.0 License by IOP Publishing Ltd and Sissa Medialab srl. Any further distribution of this work must maintain attribution to the author(s) and the published article's title, journal citation and DOI.
Appears in Collections:Published Articles, Dept. of Physics and Astronomy

Files in This Item:
File Description SizeFormat 
jinst16_09_p09009.pdfPublished (publisher PDF)1.12 MBAdobe PDFView/Open


Items in LRA are protected by copyright, with all rights reserved, unless otherwise indicated.