Please use this identifier to cite or link to this item: http://hdl.handle.net/2381/39823
Title: Is dynamic cerebral autoregulation measurement using transcranial Doppler ultrasound reproducible in the presence of high concentration oxygen and carbon dioxide?
Authors: Minhas, Jatinder S.
Syed, Nazia F.
Haunton, Victoria J.
Panerai, Ronney B.
Robinson, Thompson G.
Mistri, Amit K.
First Published: 18-Apr-2016
Publisher: IOP Publishing
Citation: Physiological Measurement, 37 (5), pp. 673-682
Abstract: Reliability of cerebral blood flow velocity (CBFV) and dynamic cerebral autoregulation estimates (expressed as autoregulation index: ARI) using spontaneous fluctuations in blood pressure (BP) has been demonstrated. However, reliability during co-administration of O2 and CO2 is unknown. Bilateral CBFV (using transcranial Doppler), BP and RR interval recordings were performed in healthy volunteers (seven males, four females, age: 54  ±  10 years) on two occasions over 9  ±  4 d. Four 5 min recordings were made whilst breathing air (A), then 5%CO2 (C), 80%O2 (O) and mixed O2  +  CO2 (M), in random order. CBFV was recorded; ARI was calculated using transfer function analysis. Precision was quantified as within-visit standard error of measurement (SEM) and the coefficient of variation (CV). CBFV and ARI estimates with A (SEM: 3.85 & 0.87; CV: 7.5% & 17.8%, respectively) were comparable to a previous reproducibility study. The SEM and CV with C and O were similar, though higher values were noted with M; Bland-Altman plots indicated no significant bias across all gases for CBFV and ARI (bias  <0.06 cm s(-1) and  <0.05, respectively). Thus, transcranial-Doppler-ultrasound-estimated CBFV and ARI during inhalation of O2 and CO2 have acceptable levels of reproducibility and can be used to study the effect of these gases on cerebral haemodynamics.
DOI Link: 10.1088/0967-3334/37/5/673
ISSN: 0967-3334
eISSN: 1361-6579
Links: http://iopscience.iop.org/article/10.1088/0967-3334/37/5/673/meta;jsessionid=948D1C30E9AA3753FFDC970BA69EEAFB.ip-10-40-2-120
http://hdl.handle.net/2381/39823
Version: Post-print
Status: Peer-reviewed
Type: Journal Article
Rights: Copyright © 2016, IOP Publishing. Deposited with reference to the publisher’s archiving policy available on the SHERPA/RoMEO website.
Appears in Collections:Published Articles, Dept. of Cardiovascular Sciences

Files in This Item:
File Description SizeFormat 
PhysM_Rev 1 JM.pdfPost-review (final submitted author manuscript)2.08 MBAdobe PDFView/Open


Items in LRA are protected by copyright, with all rights reserved, unless otherwise indicated.