Please use this identifier to cite or link to this item:
Title: Novel electronics for MCP imaging detectors
Authors: Lees, J. E.
Bassford, D. J.
Whitford, C.
Blake, O.
First Published: 2005
Citation: Nuclear Instruments and Methods in Physics Research A, 2005, 537 (03-Jan), pp.621-625
Abstract: We describe novel electronics for processing signals from a microchannel plate detector using a resistive anode readout. High sampling-rate (MHz) Analogue-to-Digital Converters (ADCs) coupled to a Digital Signal Processor (DSP) combine to offer high spatial resolution 70 micrometre low noise <0.1 count cm^-1 s^-1 imaging over a large 93 mm x 93 mm imaging area. These new low-cost electronics form an integral part of a MCP-based system for digital autoradiography designed to image radiolabelled thin tissue slices and 1- and 2-D electrophoresis gels.
DOI Link: 10.1016/j.nima.2004.08.067
ISSN: 0168-9002
Type: Article
Appears in Collections:Published Articles, Dept. of Physics and Astronomy